DQRID : D060227.1
Start DateStart TimeEnd DateEnd Time
02/24/2006060002/27/20060700
more
Subject:
TWP/MPL/C3 - Backscatter profiles are inaccurate due to computer HD corruption.
DataStreams:twpmplC3.00, twpmplC3.a1
Description:
The data shows a consistent cloud deck at 7.5km with what appears to be a duplication of 
the data from ground zero. The temperature and energy monitor values are also affected. 
The computer was found with multiple corruption problems. These corruptions may have cause 
the ghost data at 7.5km.  A chkdsk was performed which appears to have corrected the 
problem.
Suggestions: 
No correction or work around is possible.
Measurements:twpmplC3.a1:
  • Laser Temperature(laser_temp)
  • Preliminary cloud base height above ground level(preliminary_cbh)
  • Laser output energy per pulse(energy_monitor)
  • Background signal(background_signal)
  • Instrument Temperature(instrument_temp)
  • Attenuated backscatter(detector_counts)
  • Detector Temperature(detector_temp)
more
twpmplC3.00:
  • (Raw data stream - documentation not supported)
more