Subject: | NIM/MWRP/M1 - Instrument noise problem |
DataStreams: | nimmwrpM1.00, nimmwrprawM1.00, nimmwrpM1.b1
|
Description: | There are spikes and elevated noise in MWRP data. The origine of the spikes is RF
interference from various sources. All brightness temperatures are affected, but in particular
the 5 K-band channels. LWP retrievals are noisy and affected by spikes as a result. |
Suggestions: | null |
Measurements: | nimmwrpM1.b1: - Retrieved water vapor density(waterVaporDensity)
- Derived relative humidity(relativeHumidity)
- Interpolated dewpoint temperature(dewpointTemperature)
- Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
- Derived virtual temperature(virtualTemperature)
- Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
- Retrieved liquid water path(liquidWaterPath)
- Temperature used for calculation(temperature)
- Interpolated water vapor mixing ratio(waterVaporMixingRatio)
- Retrieved cloud liquid water content(liquidWaterContent)
- Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
- Microwave brightness temperature(brightnessTemperature)
more
nimmwrprawM1.00: - (Raw data stream - documentation not supported)
more
nimmwrpM1.00: - (Raw data stream - documentation not supported)
more
|