PYE/GNDRAD/M1 - Reprocessed: IRT configuration and calibration
DataStreams:pyegndrad60sM1.b1, pyegndrad20sM1.a0
Description:
The configuration of the IRT was inadvertently set to default and the program of the data 
loggers were changed to that used at TWP.C3.  This caused the calibration of the data to 
be wrong. It was corrected by reconfiguring the IRT and changing the conversion factors 
in the data logger program.

These data have been reprocessed to apply correct calibrations and the reprocessed data 
were archived 20070212.
Suggestions: 
Retrieve reprocessed data from ARM Archive.