Subject: | NIM/MWR/M1 - Reprocessed: Recalibration to correct for occasional overheating. |
DataStreams: | nimmwrlosM1.b1, nimmwrtipM1.a1
|
Description: | The radiometer was intermittently thermally unstable resulting in poor calibrations for
four brief time periods. These data have been reprocessed to apply corrected calibrations.
The affected time periods were:
20060302-20060303
20060423-20060425
20060512-20060515
20060531-20060603 |
Suggestions: | Retrieve the reprocessed data from the ARM Archive. |
Measurements: | nimmwrtipM1.a1: - 23.8 GHz Blackbody signal(bb23)
- 23.8 GHz blackbody+noise injection signal(bbn23)
- Temperature correction coefficient at 31.4 GHz(tc31)
- Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
- 23.8 GHz goodness-of-fit coefficient(r23)
- Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
- Total water vapor along zenith path using tip-derived brightness temperatures(vaptip)
- 31.4 GHz sky signal(tipsky31)
- 23.8 GHz sky brightness temperature derived from tip curve(tbsky23tip)
- (tknd)
- Temperature correction coefficient at 23.8 GHz(tc23)
- Correlation coefficients for tip curve at 31.4 GHz(r31)
- Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
- 31.4 GHz blac2body+noise injection signal(bbn31)
- 31.4 GHz blackbody(bb31)
- Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
- 31.4 GHz sky brightness temperature derived from tip curve(tbsky31tip)
- Total liquid water along zenith path using tip-derived brightness temperatures(liqtip)
- Blackbody kinetic temperature(tkbb)
- Mixer kinetic (physical) temperature(tkxc)
- 23.8 GHz sky signal(tipsky23)
more
nimmwrlosM1.b1: - Blackbody kinetic temperature(tkbb)
- Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
- 31.4 GHz blac2body+noise injection signal(bbn31)
- Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
- 23.8 GHz Blackbody signal(bb23)
- 23.8 GHz sky signal(sky23)
- Averaged total liquid water along LOS path(liq)
- Temperature correction coefficient at 23.8 GHz(tc23)
- (tknd)
- Mean 23.8 GHz sky brightness temperature(tbsky23)
- Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
- Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
- Mixer kinetic (physical) temperature(tkxc)
- Total water vapor along LOS path(vap)
- 23.8 GHz blackbody+noise injection signal(bbn23)
- 31.4 GHz blackbody(bb31)
- 31.4 GHz sky signal(sky31)
- Temperature correction coefficient at 31.4 GHz(tc31)
- Mean 31.4 GHz sky brightness temperature(tbsky31)
more
|