DQRID : D060808.1
Start DateStart TimeEnd DateEnd Time
08/15/2006190008/18/20060800
more
Subject:
SGP/MWR/B6 - Reprocessed: Corrected calibration
DataStreams:sgpmwrtipB6.a1, sgpmwrlosB6.b1
Description:
The instrument was overheating so the radiometer temperature set point was increased from 
56.3 C to 57.8 C.  The data have been reprocessed with corrected calibration 
coefficients.  Reprocessed MWRLOS data were archived January 2007.  MWRTIP data will not be 
reprocessed.
Suggestions: 
Retrieve the reprocessed data for use in your analysis.
Measurements:sgpmwrtipB6.a1:
  • (tknd)
  • Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
  • Mixer kinetic (physical) temperature(tkxc)
  • 31.4 GHz sky signal(tipsky31)
  • Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
  • Noise injection temp at 31.4 GHz derived from this tip(tnd31I)
  • Total water vapor along zenith path using tip-derived brightness temperatures(vaptip)
  • 23.8 GHz sky brightness temperature derived from tip curve(tbskytip23)
  • Temperature correction coefficient at 23.8 GHz(tc23)
  • Correlation coefficients for tip curve at 31.4 GHz(r31)
  • 23.8 GHz goodness-of-fit coefficient(r23)
  • Temperature correction coefficient at 31.4 GHz(tc31)
  • Total liquid water along zenith path using tip-derived brightness temperatures(liqtip)
  • Noise injection temp at 23.8 GHz derived from this tip(tnd23I)
  • Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
  • 31.8 GHz sky brightness temperature derived from tip curve(tbskytip31)
  • Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
  • 23.8 GHz sky signal(tipsky23)
more
sgpmwrlosB6.b1:
  • Mixer kinetic (physical) temperature(tkxc)
  • Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
  • Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
  • Mean 31.4 GHz sky brightness temperature(tbsky31)
  • Temperature correction coefficient at 31.4 GHz(tc31)
  • Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
  • Total water vapor along LOS path(vap)
  • Averaged total liquid water along LOS path(liq)
  • Mean 23.8 GHz sky brightness temperature(tbsky23)
  • Temperature correction coefficient at 23.8 GHz(tc23)
  • (tknd)
  • Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
more

Close this window