DQRID : D061117.1
Start DateStart TimeEnd DateEnd Time
10/01/2006000011/17/20061713
more
Subject:
NIM/MWRP/M1 - Reprocessed: K-band calibration drift corrected
DataStreams:nimmwrpM1.b1
Description:
During October and November there was a drift in the K-Band calibration of the MWRP. This 
resulted in brightness temperatures that were too high.

New calibration coefficients were derived from the median value of the noise diode 
temperature for the months of October and November and the data were reprocessed to apply the 
corrected calibrations.  The reprocessed data were archived in January 2007.
Suggestions: 
Retrieve the reprocessed data for use in your analysis.
Measurements:nimmwrpM1.b1:
  • Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
  • Retrieved liquid water path(liquidWaterPath)
  • Retrieved cloud liquid water content(liquidWaterContent)
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
  • Microwave brightness temperature(brightnessTemperature)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)