Subject: | SGP/SIRS/E7 - Instrument noise |
DataStreams: | sgpsirsE7.b1, sgpsirs20sE7.a0
|
Description: | Beginning around 10/31, the E7 UIR experienced an increasingly noisy signal. The
thermopile signal was noisy, but case and dome thermistors are well behaved. Maintenance
reseated the UIR cable connections on 12/12/2006 which effectively eliminated the UIR noise.
The thermopile data from the upwelling pyrgeometer is noisy with resulting irradiance
variations of up to 5 W/m2, but doesn't appear to have a bias. |
Suggestions: | Use data during this time range at users discretion. |
Measurements: | sgpsirs20sE7.a0: - Instantaneous Upwelling Pyrgeometer Dome Thermistor Resistance Pyrgeometer(inst_up_long_dome_resist)
- Instantaneous Upwelling Pyrgeometer Thermopile Pyrgeometer(inst_up_long_hemisp_tp)
- Instantaneous Upwelling Pyrgeometer Case Thermistor Resistance, Pyrgeometer(inst_up_long_case_resist)
more
sgpsirsE7.b1: - Instantaneous upwelling pyrgeometer case thermistor temperature, pyrgeometer(inst_up_long_case_temp)
- Instantaneous Dome Temperature(inst_up_long_dome_temp)
- Instantaneous Upwelling Longwave Hemispheric Irradiance, Pyrgeometer(up_long_netir)
- 10 meter Upwelling Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)
- Upwelling Longwave Hemispheric Irradiance, Pyrgeometer, Minima(up_long_hemisp_min)
- Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
Standard Deviation(up_long_hemisp_std) - Upwelling longwave hemispheric irradiance, pyrgeometer, maxima(up_long_hemisp_max)
more
|