DQRID : D070223.2
Start DateStart TimeEnd DateEnd Time
02/22/2007163003/06/20072100
more
Subject:
SGP/EBBR/E8 - Failed Temp/RH sensor
DataStreams:sgp30ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.b1
Description:
The right Temperature-Humidity sensor failed. The measured temperature gradients were 
incorrect and thus the sensible and latent heat fluxes were incorrect for this extended 
period.

The sensor was replaced March 6.
Suggestions: 
Measurements:sgp5ebbrE8.b1:
  • Top vapor pressure(vapor_pressure_top)
  • Bottom vapor pressure(vapor_pressure_bottom)
  • Bottom T/RH sensor temperature(temp_trh_bottom)
  • Top T/RH sensor temperature(temp_trh_top)
more
sgp15ebbrE8.b1:
  • Right T/RH sensor temperature (resistance ratio)(temp_trh_right_rr)
more
sgp30ebbrE8.b1:
  • Top vapor pressure(vapor_pressure_top)
  • Bottom T/RH sensor temperature(temp_trh_bottom)
  • Bottom vapor pressure(vapor_pressure_bottom)
  • Surface latent heat flux(latent_heat_flux)
  • Bowen ratio(bowen_ratio)
  • Top T/RH sensor temperature(temp_trh_top)
  • Surface sensible heat flux(sensible_heat_flux)
more

Close this window