DQRID : D070406.6
Start DateStart TimeEnd DateEnd Time
02/04/2007231702/17/20070400
more
Subject:
TWP/GNDRAD/C2 - Reprocessing Notice (RID 362)
DataStreams:twpgndrad60sC2.b1
Description:
See below for details of reprocessing that was completed on 20150820.  Original DQR 
category: 4 (Red, Incorrect).

-----------------------------------------------------------

The configuration of the IRT was set to that used at TWP/C3. This caused the calibration 
of the data to be wrong. It was corrected by reconfiguring the IRT to match the datalogger 
program.
   
These data can be reprocessed and corrected by adding 10K, i.e. sfc_ir_temp(new) = 
sfc_ir_temp(old) + 10K.
Suggestions: 
Measurements:twpgndrad60sC2.b1:
  • Surface Infra-Red Temperature, Maxima(sfc_ir_temp_max)
  • Surface Infra-Red Temperature, Minima(sfc_ir_temp_min)
  • Surface infra-red temperature(sfc_ir_temp)