DQRID : D070618.2
Start DateStart TimeEnd DateEnd Time
04/16/2007000006/18/20072000
more
Subject:
SGP/SMOS/E13 - RH sensor failure
DataStreams:sgp30smosE13.b1, sgp1440smosE13.b1, sgp1smosE13.b1
Description:
The T/RH sensor relative humidity data is suspect during this time.  Values from this 
sensor failed the 6-month check and subsequent bi-weekly checks.  The RH values were lower by 
larger than the manufacturer suggested uncertainty of 3%.
Suggestions: 
Use RH data from nearby sensors such as the EBBR and/or THWAPS.
Measurements:sgp1440smosE13.b1:
  • Time of maximum relative humidity(time_max_rh)
  • Minimum Relative Humidity(min_rh)
  • Time of maximum vapor pressure(time_max_vap_pres)
  • Time of Relative Humidity(time_min_rh)
  • Daily maximum relative humidity(max_rh)
  • Time of minimum vapor pressure(time_min_vap_pres)
  • Daily minimum vapor pressure(min_vap_pres)
  • Maximum Vapor Pressure(max_vap_pres)
more
sgp30smosE13.b1:
  • Standard Deviation of Relative Humidity(sd_rh)
  • Standard deviation of vapor pressure(sd_vap_pres)
  • Relative Humidity (hybrid-b grid)(rh)
  • Vapor Pressure (kiloPascals)(vap_pres)
more
sgp1smosE13.b1:
  • Relative Humidity (hybrid-b grid)(rh)
  • Vapor Pressure (kiloPascals)(vap_pres)
more