DQRID : D070917.1
Start DateStart TimeEnd DateEnd Time
09/02/2006040009/14/20061900
more
Subject:
SGP/EBBR/E13 - Radiometer dome broken
DataStreams:sgp5ebbrE13.b1, sgp30ebbrE13.b1, sgp15ebbrE13.b1
Description:
The top net radiometer dome was broken, yielding net radiation, sensible heat flux and 
latent heat flux that were too low.

The BEFLUX (SIRS) net radiation can be used as a replacement for the EBBR net radiation.

To correct the sensible and latent heat fluxes, determine the ratio of SIRS net radiation 
to EBBR net radiation for each daylight half hour and multiply that ratio times H and LE 
to obtain approximate corrected flux values.
Suggestions: 
Correct fluxes as directed.
Measurements:sgp15ebbrE13.b1:
  • Net radiation (mV)(net_radiation_mv)
more
sgp5ebbrE13.b1:
  • Net radiation(net_radiation)
more
sgp30ebbrE13.b1:
  • Surface sensible heat flux(sensible_heat_flux)
  • Surface latent heat flux(latent_heat_flux)
  • Net radiation(net_radiation)
more