DQRID : D071016.1
Start DateStart TimeEnd DateEnd Time
09/17/2007150010/09/20071840
more
Subject:
FKB/MWRP/M1 - Reprocessed: Updated 30-GHz calibration coefficient
DataStreams:fkbmwrpM1.b1
Description:
On 9/17 the instrument was calibrated with liquid nitrogen. For the 30-GHz channel the 
results of calibration were not optimal and were causing a negative bias in the retrieved 
LWP.

On 10/9 the 30-GHz calibration parameters were updated with values derived from tip 
curves.

Data between 9/17 and 10/9 were reprocessed to correct the 30-GHz calibration.  
Reprocessed data were archived 20080122.
Suggestions: 
Retrieve reprocessed data for your analysis
Measurements:fkbmwrpM1.b1:
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
  • Microwave brightness temperature(brightnessTemperature)
  • Retrieved liquid water path(liquidWaterPath)
more

Close this window