Subject: | TWP/GNDRAD/C2 - Reprocessed: Incorrect calibration |
DataStreams: | twpgndrad60sC2.b1, twpgndrad20sC2.a0
|
Description: | The Gndrad PSP was changed out 11/25 but the datalogger program was not updated until
11/30.
PSP #31293 installed 20071125.2202. Correct calibration coefficient = 123.46.
Reprocessed to correct upwelling shortwave hemispheric irradiance (US) by:
US(new) = (US(old) / 123.56 ) * 123.46
Reprocessing completed March 2008. |
Suggestions: | Retrieve reprocessed data for your analysis. |
Measurements: | twpgndrad60sC2.b1: - Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Maxima(up_short_hemisp_max)
- 25 meter Upwelling Shortwave Hemispheric Irradiance, Pyranometer(up_short_hemisp)
- Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer,
Standard Deviation(up_short_hemisp_std) - Upwelling shortwave hemispheric irradiance, pyranometer, minima(up_short_hemisp_min)
more
twpgndrad20sC2.a0: - Instantaneous Upwelling Shortwave Hemispheric Irradiance, Pyranometer(inst_up_short_hemisp)
more
|