DQRID : D071217.6
Start DateStart TimeEnd DateEnd Time
07/12/2005153003/28/20081722
more
Subject:
SGP/IRT/E15 - Instrument noise problem
DataStreams:sgpirtE15.b1, sgpirtE15.00, sgpirt200msE15.a1
Description:
The data was much noisier than other nearby sites (a factor of 4-5 times more). The 
connections and communication and other settings were checked. The IRT was replaced on 
1/29/2008 at 17:37 but the response time was incorrectly set to 0.01 s. This problem was 
corrected when the setting was changed to 0.10 s.
Suggestions: 
Measurements:sgpirtE15.00:
  • (Raw data stream - documentation not supported)
more
sgpirtE15.b1:
  • Sky Infra-Red Temperature(sky_ir_temp)
  • Sky Infra-Red Temperature Minima(sky_ir_temp_min)
more
sgpirt200msE15.a1:
  • Sky Infra-Red Temperature(sky_ir_temp)
more