Subject: | NSA/MWRP/C1 - Radiometer thermally unstable |
DataStreams: | nsamwrpC1.b1
|
Description: | Between Dec 21 and Dec 23 the external temperature dropped to - 38 C. This caused a
problem with the stability of the noise diode mount temperature and consequent problem with
calibration. |
Suggestions: | Avoid data between 12/21 and 12/23. |
Measurements: | nsamwrpC1.b1: - Derived relative humidity(relativeHumidity)
- Retrieved cloud liquid water content(liquidWaterContent)
- Interpolated water vapor mixing ratio(waterVaporMixingRatio)
- Derived virtual temperature(virtualTemperature)
- Microwave brightness temperature(brightnessTemperature)
- Interpolated dewpoint temperature(dewpointTemperature)
- Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
- Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
- Retrieved water vapor density(waterVaporDensity)
- Retrieved liquid water path(liquidWaterPath)
- Temperature used for calculation(temperature)
- Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
more
|