NSA/MWRP/C1 - Reprocessed: LN2 calibration
DataStreams:nsamwrpC1.b1
Description:
On 4/9 a spare radiometer was installed at the site. On 4/18 a LN2 calibration was 
conducted. Upon examination of the data it appeared that the LN2 calibration was generally 
successful, however the 30-GHz channel was not calibrated properly.
New calibration coefficients for all K-band channels were derived and data for the whole 
month should be reprocessed with these newly derived coefficients. I will provide the new 
configuration file.

An explanation of the effects of the poor calibration is given in the Instrument Mentor 
Monthly Summary report. The poor calibration affected the LWP in particular. Re-calibration 
will correct the problem.

Data reprocessed 20080610.
Suggestions: 
Use reprocessed data
Measurements:nsamwrpC1.b1:
  • Derived relative humidity(relativeHumidity)
  • Retrieved cloud liquid water content(liquidWaterContent)
  • Interpolated water vapor mixing ratio(waterVaporMixingRatio)
  • Microwave brightness temperature(brightnessTemperature)
  • Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
  • Retrieved liquid water path(liquidWaterPath)
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)