DQRID : D080523.5
Start DateStart TimeEnd DateEnd Time
04/09/2008153905/01/20080300
more
Subject:
NSA/MWRP/C1 - Increased RF intereference
DataStreams:nsamwrpC1.b1
Description:
The spare radiometer experienced increased levels of RF interference. The interference is 
prominent on the 30-GHz channel and LWP retrievals that appear particularly noisy.
Suggestions: 
The user can smooth the data with some running average to mitigate the noise
Measurements:nsamwrpC1.b1:
  • Microwave brightness temperature(brightnessTemperature)
  • Retrieved liquid water path(liquidWaterPath)
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
more