Subject:
NSA/METTWR/C1 - CMH required Calibration
DataStreams:nsamettwr4hC1.b1
Description:
The Chilled Mirror was showing some signs of being out of calibration.  Temperatures were 
at times higher than co-located temp probes and the dew point temperature values at times 
were not consistent with co-located RH sensors.
Suggestions: 
Use T/RH probe data at 2m.
Measurements:nsamettwr4hC1.b1:
  • Chilled Mirror Dew Point(CMHDP)
  • Chilled Mirror Calculated Saturation Vapor Pressure(SatVPCMH)
  • Chilled Mirror Calculated Vapor Pressure(VPCMH)
  • Chilled Mirror Temperature(CMHTemp)
  • Chilled Mirror Calculated Relative Humidity(CMHRH)
more