DQRID : D080710.2
Start DateStart TimeEnd DateEnd Time
07/01/2008000007/18/20081950
more
Subject:
NSA/MWRP/C1 - Reprocessed: LN2 calibrations applied
DataStreams:nsamwrpC1.b1
Description:
The V-band measurements were incorrect. The quantities affected were the temperature and 
humidity profiles.

The instrument was re-calibrated on July 18.  The new calibrations have been applied to 
the data for this time period.  The reprocessed data were archived in January 2009.
Suggestions: 
Measurements:nsamwrpC1.b1:
  • Derived relative humidity(relativeHumidity)
  • Retrieved cloud liquid water content(liquidWaterContent)
  • Interpolated water vapor mixing ratio(waterVaporMixingRatio)
  • Derived virtual temperature(virtualTemperature)
  • Microwave brightness temperature(brightnessTemperature)
  • Interpolated dewpoint temperature(dewpointTemperature)
  • Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
  • Retrieved water vapor density(waterVaporDensity)
  • Retrieved liquid water path(liquidWaterPath)
  • Temperature used for calculation(temperature)
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
more