Subject:
HFE/MWRHF/M1 - Frosting/Condensation
DataStreams:hfemwrhfM1.b1
Description:
The heater module was installed on 6/4/2008. Before this time there is condensation on the 
window during night time and early morning. On 8/18 the heater appears to have failed 
and nighttime condensation re-occurred on the radome during conditions of high humidity.

Users can identify the segments affected by observing a smooth steady increase in the 
brightness temperature (BT). BT values can increase to more than 250 K and then smoothly 
decrease. The RH during these events is usually above 70%. Users should disregard data 
affected by condensation.