GRW/MWR/M1 - Loose mirror
DataStreams:grwmwrlosM1.b1, grwmwrtipM1.a1
Description:
The data from the MWR were suspicious and became progressively worse. The problem was 
traced back to the fact that the screw that keeps the mirror in place had become loose and 
the mirror was freely rotating. 

The mirror was tightened on 6/9/2009, however the measurements were noisy until 6/16, when 
the instrument was rebooted and all the connections checked. 

Measurements between 5/1 and 6/9 should not be used. Data between 6/9 and 6/16 are usable, 
although they are very noisy.
Suggestions: 
Measurements:grwmwrtipM1.a1:
  • 23.8 GHz Blackbody signal(bb23)
  • Total liquid water along zenith path using tip-derived brightness temperatures(liqtip)
  • 31.4 GHz sky signal(tipsky31)
  • 31.4 GHz sky brightness temperature derived from tip curve(tbsky31tip)
  • 23.8 GHz blackbody+noise injection signal(bbn23)
  • 31.4 GHz blac2body+noise injection signal(bbn31)
  • 23.8 GHz sky signal(tipsky23)
  • 23.8 GHz sky brightness temperature derived from tip curve(tbsky23tip)
  • 31.4 GHz blackbody(bb31)
  • Total water vapor along zenith path using tip-derived brightness temperatures(vaptip)
more
grwmwrlosM1.b1:
  • Mean 31.4 GHz sky brightness temperature(tbsky31)
  • Averaged total liquid water along LOS path(liq)
  • Mean 23.8 GHz sky brightness temperature(tbsky23)
  • 23.8 GHz sky signal(sky23)
  • 23.8 GHz Blackbody signal(bb23)
  • 31.4 GHz sky signal(sky31)
  • 31.4 GHz blac2body+noise injection signal(bbn31)
  • 31.4 GHz blackbody(bb31)
  • Total water vapor along LOS path(vap)
  • 23.8 GHz blackbody+noise injection signal(bbn23)
more

Close this window