Subject:
SGP/AOS/C1 - Suspected leak in CCN and or problem with instrument optics
DataStreams:sgpaosccnC1.a1
Description:
The CCN signal started a slow decline during this period. The instrument was removed and 
shipped back to the mentor for repair. A leak was found in the instrument which would've 
decreased the over all signal. Whether this leak affected the rate of CCN growth with %SS 
is uncertain.
Suggestions: 
Don't use the data as the instrument wasn't operating properly.
Measurements:sgpaosccnC1.a1:
  • AOS CCN volumetric flowrate of sheath air(CCN_Q_sheath)
  • Cloud condensation nuclei number density(N_CCN)
  • AOS CCN number concentration by bin(N_CCN_dN)
  • AOS CCN sample saturation setpoint value reported by instrument(CCN_ss_set)
  • AOS CCN sample supersaturation calculated by model)(CCN_ss_calc)
  • AOS CCN volumetric flowrate of sample air(CCN_Q_sample)
  • Bin number of lowest channel of OPC include in summation of N_CCN(N_CCN_bin_number)