Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/19/2009 | 1844 | 01/15/2010 | 1820 |
Subject: | GRW/AOS/M1 - CPC TEC failure |
DataStreams: | grwaosM1.a1 |
Description: | The thermal electric cooler, which controls the temperature in the Condensation Particle Counter, failed. A new TEC was mailed to GRW. The TEC was replaced and the CPC was working again on Jan. 15th, 2010. |
Suggestions: | CPC data are missing and unrecoverable. |
Measurements: | grwaosM1.a1: more
|