Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 10/31/2009 | 2359 |
Subject: | SGP/VCEIL/C1/B4/B5/B6 - Reprocess: Incorrect serial_number and sum_backscatter fields |
DataStreams: | sgpvceil25kC1.b1, sgpvceil25kB5.b1, sgpvceil25kB6.b1, sgpvceil25kB4.b1 |
Description: | Attribute serial_number: In most cases, this field is not representative of the actual serial number of the instrument because the value did not change when an instrument changed locations or was replaced with a spare. Variable sum_backscatter: An out-of-range condition for the "sum of detected and normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 0.0999 /srad. This DQR also applies to a1 level data collected prior to b1 level. |
Suggestions: | |
Measurements: | sgpvceil25kC1.b1: more
sgpvceil25kB5.b1: more sgpvceil25kB6.b1: more sgpvceil25kB4.b1: more |