DQRID : D100521.3
Start DateStart TimeEnd DateEnd Time
03/26/2010000005/04/20101935
more
Subject:
SGP/SIRS/E21 - Tracker failure
DataStreams:sgpsirsE21.b1, sgpsirs20sE21.a0
Description:
First apparent on 03/26, the short_direct_normal values were incorrect after approximately 
1700GMT each day. The tracker was found to have excessive clock drift and was reset by 
site ops on 05/04/2010 and returned to service at 1935GMT. The down_short_diffuse_hemisp 
values appear unaffected by the tracker misalignment.

The short_direct_normal data values are suspect after approximately 1800GMT each day from 
03/26 to 04/26 and after 1700GMT from 04/27 to the end of the period.  The 
short_direct_normal data are not recoverable but can be calculated indirectly using the following 
relationship for each minute of data:
down_short_hemsip = short_direct_normal * cos(zenith angle) + down_short_diffuse
Suggestions: 
Measurements:sgpsirs20sE21.a0:
  • Instantaneous direct normal shortwave irradiance, pyrheliometer thermopile
    voltage(inst_direct_normal)
more
sgpsirsE21.b1:
  • Shortwave direct normal irradiance, pyrheliometer(short_direct_normal)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Standard Deviation(short_direct_normal_std)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • Shortwave direct normal irradiance, pyrheliometer, minima(short_direct_normal_min)
more