DQRID : D100909.1
Start DateStart TimeEnd DateEnd Time
07/01/2010000009/08/20101532
more
Subject:
GRW/MWRP/M1 - Reprocessed: Updated calibration coefficients
DataStreams:grwmwrpM1.b1
Description:
On 9/8/2010 the calibration of of the K-band channels was updated. This caused a slight 
decrease of the retrieved liquid water path. 

Updated calibrations have been applied to data from 7/1-9/8.  The LWP was decreased 
slightly as a result of the recalibration.  Reprocessed data were archived in November 2010.
Suggestions: 
Measurements:grwmwrpM1.b1:
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
  • Retrieved cloud liquid water content(liquidWaterContent)
  • Microwave brightness temperature(brightnessTemperature)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
  • Retrieved liquid water path(liquidWaterPath)
  • Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
more