DQRID : D101215.3
Start DateStart TimeEnd DateEnd Time
08/09/2010153410/04/20101445
more
Subject:
SGP/IRT/E13 - Instrument output invalid
DataStreams:sgpirt200msE13.a1, sgpirtE13.b1
Description:
The signal was missing for 25 min and then changed to a constant values of 
sky_ir_temp=262.22K and ref_ir_temp=313.10K. The problem was corrected when the intrument was 
power-cycled.
Suggestions: 
Measurements:sgpirtE13.b1:
  • Sky Infra Red Temperature Maxima(sky_ir_temp_max)
  • Internal Reference Infra Red Temperature(ref_ir_temp)
  • Sky Infra-Red Temperature(sky_ir_temp)
  • Sky Infra-Red Temperature Minima(sky_ir_temp_min)
  • Sky infrared temperature, standard deviation(sky_ir_temp_std)
  • ref_ir_temp_std
more
sgpirt200msE13.a1:
  • Internal Reference Infra Red Temperature(ref_ir_temp)
  • Sky Infra-Red Temperature(sky_ir_temp)
more