DQRID : D110210.1
Start DateStart TimeEnd DateEnd Time
11/01/2010000001/19/20111400
more
Subject:
NSA/GVRP/C1 - Reprocessed: Calibration drift
DataStreams:nsagvrpC1.b1
Description:
The most transparent channels (170-176 GHz) exhibited signs of a drift in the calibration. 
The radiometer was calibrated on Jan 17 2011.  Data for 20101101-20110117 were 
recalibrated and reprocessed.  The reprocessed data were archived in May 2011.  

After the calibration there was a problem restarting the software. The software was 
successfully restarted on Jan 19.