Description: | Apparent lightning strike in the instrument cluster area resulted in a failed NIP. The
SIRS was due for calibration change outs so the US, DS, DD and NIP were replaced. The
updated program was loaded at approximately 1730 GMT.
An estimate of direct normal values can be derived using the following formula:
short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle) |
Measurements: | sgpsirsE13.b1: - Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
- Shortwave direct normal irradiance, pyrheliometer(short_direct_normal)
- Shortwave direct normal irradiance, pyrheliometer, minima(short_direct_normal_min)
- Shortwave Direct Normal Irradiance, Pyrgeometer, Standard Deviation(short_direct_normal_std)
more
sgpsirs20sE13.a0: - Instantaneous direct normal shortwave irradiance, pyrheliometer thermopile
voltage(inst_direct_normal) more
|