Subject: | PGH/MWR/M1 - Reprocessed: Incorrect calibration |
DataStreams: | pghmwrlosM1.b1, pghmwrtipM1.a1
|
Description: | When the radiometer started to collect data it was not at its final location and had an
obstructed field of view. The radiometer was moved to its final location on 7/13/2011.
After the radiometer was moved it started to collect good data for calibration.
The MWRLOS data were reprocessed with appropriate calibrations. The reprocessed data were
archived in July 2012. |
Suggestions: | Use reprocessed MWRLOS data |
Measurements: | pghmwrtipM1.a1: - Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
- Temperature correction coefficient at 31.4 GHz(tc31)
- Noise injection temp at 23.8 GHz derived from this tip(tnd23I)
- Noise injection temp at 31.4 GHz derived from this tip(tnd31I)
- Temperature correction coefficient at 23.8 GHz(tc23)
- Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
- Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
- 31.4 GHz sky brightness temperature derived from tip curve(tbsky31tip)
- Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
- 23.8 GHz sky brightness temperature derived from tip curve(tbsky23tip)
more
pghmwrlosM1.b1: - Mean 23.8 GHz sky brightness temperature(tbsky23)
- Noise injection temp at nominal temperature at 23.8 GHz(tnd_nom23)
- Mean 31.4 GHz sky brightness temperature(tbsky31)
- Noise injection temp at nominal temperature at 31.4 GHz(tnd_nom31)
- Noise injection temp at 31.4 GHz adjusted to tkbb(tnd31)
- Noise injection temp at 23.8 GHz adjusted to tkbb(tnd23)
- Temperature correction coefficient at 23.8 GHz(tc23)
- Temperature correction coefficient at 31.4 GHz(tc31)
more
|