Subject: | GAN/GNDRAD/M1 - Incorrect up_long_hemisp |
DataStreams: | gangndrad60sM1.b1, gangndrad20sM1.a0
|
Description: | The up_long_hemisp values from the pyrgeometer were much too low, hovering near or just
above 100 W/m^2. This was particularly noticeable when comparing these values to those
derived from the surface IRT.
On 10/05/11, at 0348Z, tower was lowered for removal of PIR SN35836 and installation of
PIR SN35839; tower raised at 0444Z. Following installation of PIR SN35839, data quality
plots indicate proper up_long_hemisp values. |
Suggestions: | |
Measurements: | gangndrad20sM1.a0: - Instantaneous Upwelling Pyrgeometer Dome Thermistor Resistance, Pyrgeometer(inst_up_long_hemisp_dome_resist)
- Instantaneous Upwelling Pyrgeometer Thermopile Pyrgeometer(inst_up_long_hemisp_tp)
- Instantaneous Upwelling Pyrgeometer Case Thermistor Resistance, Pyrgeometer(inst_up_long_hemisp_case_resist)
more
gangndrad60sM1.b1: - 10 meter Upwelling Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)
- Upwelling longwave hemispheric irradiance, pyrgeometer, maxima(up_long_hemisp_max)
- Instantaneous Upwelling Pyrgeometer Case Thermistor Temperature(inst_up_long_hemisp_case_temp)
- Instantaneous Upwelling Pyrgeometer Thermopile Pyrgeometer(inst_up_long_hemisp_tp)
- Upwelling Longwave Hemispheric Irradiance, Pyrgeometer, Minima(up_long_hemisp_min)
- Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
Standard Deviation(up_long_hemisp_std) - Instantaneous Upwelling Pyrgeometer Dome Thermistor Temperature(inst_up_long_hemisp_dome_temp)
more
|