Description: | The MPL was showing signs of laser wear and was no longer correlating well with either the
local VCEIL or the local KAZR. Instrument was not reporting any cloud signatures at
night, and showed a notable lack of scatter near the surface. The deterioration appears to
have started around 7/01 and slowly progressed.
The system was replaced 8/24/2011.
Depolarization values are incorrect. However, upper level cloud heights may still be
usable. |
Measurements: | twpmplpolavgC1.s1: - Background signal standard deviation, cross-pol(background_signal_std_cross_pol)
- Background signal, cross pol data(background_signal_cross_pol)
- Attenuated backscatter, co-pol(signal_return_co_pol)
- Background signal standard deviation, copol data(background_signal_std_co_pol)
- (aqc_signal_return_co_pol)
- Background signal, co-pol(background_signal_co_pol)
- Attenuated backscatter, cross pol(signal_return_cross_pol)
- (aqc_signal_return_cross_pol)
more
twpmplpolavgC1.c1: - Background signal, cross pol data(background_signal_cross_pol)
- (aqc_signal_return_cross_pol)
- Background signal standard deviation, cross-pol(background_signal_std_cross_pol)
- Attenuated backscatter, cross pol(signal_return_cross_pol)
- (aqc_signal_return_co_pol)
- Background signal standard deviation, copol data(background_signal_std_co_pol)
- Background signal, co-pol(background_signal_co_pol)
- Attenuated backscatter, co-pol(signal_return_co_pol)
more
|