Subject: | SGP/QCRAD/E1 - Intermittently missing downwelling diffuse |
DataStreams: | sgpqcrad1longE1.c2, sgpqcrad1longE1.s2
|
Description: | Starting on or about 10/26/2005 the downwelling diffuse hemispheric irradiance (Eppley
8-48) instrument had intermittent missing values. Technicians found a loose wire on the
instrument connector and repaired it on 4/4/2006.
This means that Best Estimate Global can't be calculated from the sum of direct and
diffuse. It also means that some quality checks can't be done.
Problem documented in instrument DQR D060331.2. |
Suggestions: | null |
Measurements: | sgpqcrad1longE1.s2: - Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer(down_short_diffuse_hemisp)
more
sgpqcrad1longE1.c2: - Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer(down_short_diffuse_hemisp)
more
|