Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/21/1996 | 1200 | 04/10/1996 | 1600 |
09/13/1996 | 1200 | 10/09/1996 | 1600 |
09/14/1995 | 0300 | 10/13/1995 | 1500 |
06/12/1995 | 0000 | 07/07/1995 | 1600 |
Subject: | SGP/QCRAD/E20 - Tracker alignment issue |
DataStreams: | sgpqcrad1longE20.s2, sgpqcrad1longE20.s1, sgpqcrad1longE20.c2, sgpqcrad1longE20.c1 |
Description: | Tracker appears to be out of alignment or inoperative during these times. The short_direct_normal (NIP) and down_short_diffuse_hemisp (8-48) data appear to be bad and not recoverable. There were no instrument DQRs on this problem at the time this report was submitted. |
Suggestions: | null |
Measurements: | sgpqcrad1longE20.s1: more
sgpqcrad1longE20.c1: more sgpqcrad1longE20.s2: more sgpqcrad1longE20.c2: more |