DQRID : D120417.15
Start DateStart TimeEnd DateEnd Time
01/19/1996192604/22/19961630
more
Subject:
SGP/QCRAD/E12 - downwelling pyranometer data incorrect
DataStreams:sgpqcrad1longE12.s2, sgpqcrad1longE12.c2, sgpqcrad1longE12.c1, sgpqcrad1longE12.s1
Description:
Because of a bad "preamp cable", the data from the downwelling pyranometer
at site E12 would generally not fall below 400 W/m^2 even at night.  We don't
know exactly when this problem begins, so beware of all downwelling
pyranometer data taken before 960422.

This will also impact the best estimate GSW data when it uses the GSW pyranometer data.

Described in instrument DQR D960501.1
Suggestions: 
null
Measurements:sgpqcrad1longE12.s2:
  • Quality check results on field: Best Estimate Global Downwelling Shortwave
    Hemispheric Irradiance(BestEstimate_down_short_hemisp)
more
sgpqcrad1longE12.c2:
  • Quality check results on field: Best Estimate Global Downwelling Shortwave
    Hemispheric Irradiance(BestEstimate_down_short_hemisp)
  • Down-welling unshaded pyranometer voltage(down_short_hemisp)
more
sgpqcrad1longE12.s1:
  • Quality check results on field: Best Estimate Global Downwelling Shortwave
    Hemispheric Irradiance(BestEstimate_down_short_hemisp)
more
sgpqcrad1longE12.c1:
  • Quality check results on field: Best Estimate Global Downwelling Shortwave
    Hemispheric Irradiance(BestEstimate_down_short_hemisp)
  • Down-welling unshaded pyranometer voltage(down_short_hemisp)
more

Close this window