Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/11/2006 | 0000 | 04/13/2006 | 1522 |
Subject: | SGP/QCRAD/E19 - Ventilator failure |
DataStreams: | sgpqcrad1longE19.s1, sgpqcrad1longE19.c1, sgpqcrad1longE19.s2, sgpqcrad1longE19.c2 |
Description: | Dowenlling shortwave hemispheric instrument signal (Eppley PSP) is very erratic due to failed ventilator fan which was replaced during PM on 04/13/2006. Discussed in instrument DQR D060403.3. |
Suggestions: | null |
Measurements: | sgpqcrad1longE19.c1: more
sgpqcrad1longE19.s1: more sgpqcrad1longE19.c2: more sgpqcrad1longE19.s2: more |