Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/28/2003 | 0000 | 02/12/2003 | 2359 |
06/03/2004 | 0100 | 06/11/2004 | 0400 |
Subject: | TWP/QCRAD/C3 - MFRSR shading problem |
DataStreams: | twpqcrad1longC3.s1, twpqcrad1longC3.c1, twpqcrad1longC3.s2, twpqcrad1longC3.c2 |
Description: | MFRSR diffuse and direct fluxes incorrect during this time because instrument was not shading properly. Some tests on best estimate global shortwave cannot be applied. BEGSW could be incorrect if MFRSR data used in tests or fits. Documented in instrument DQR D030317.2, D050106.2 |
Suggestions: | null |
Measurements: | twpqcrad1longC3.s2: more
twpqcrad1longC3.c2: more twpqcrad1longC3.s1: more twpqcrad1longC3.c1: more |