DQRID : D120521.1
Start DateStart TimeEnd DateEnd Time
04/13/2012182204/25/20121608
more
Subject:
NSA/MWRP/C1 - Reprocessed: Incorrect calibration
DataStreams:nsamwrpC1.b1
Description:
The LN2 calibration was only partially successful, some the K-band channels did not 
calibrate properly. On 4/25 new coefficients were derived from tip curves and were uploaded. 

The data for 4/13-4/25 were reprocessed to apply the 4/25 calibrations.  Reprocessed data 
were archived in July 2012.
Suggestions: 
Measurements:nsamwrpC1.b1:
  • Retrieved cloud liquid water content(liquidWaterContent)
  • Interpolated water vapor mixing ratio(waterVaporMixingRatio)
  • Derived virtual temperature(virtualTemperature)
  • Microwave brightness temperature(brightnessTemperature)
  • Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
  • Retrieved water vapor density(waterVaporDensity)
  • Retrieved liquid water path(liquidWaterPath)
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
more