Subject:
SGP/ACSM/C1 - Water in inlet sample line
DataStreams:sgpaosacsmC1.b1, sgpaosacsmC1.00, sgpaosacsmC1.01, sgpaosacsmC1.a1
Description:
Data became progressively more noisy causing large spikes in data. These spikes started 
occurring for longer periods of time during the day before the instrument went down 
completely on 7/12.

The afternoon data (usually 15:00-24:00 UTC) was incorrect because of water condensation 
in the inlet line and aerosol collection efficiency change. The organic mass loading is 
around 0 when the condensed water was absorbing all the particles.
Suggestions: 
Do not use data
Measurements:sgpaosacsmC1.01:
  • (Raw data stream - documentation not supported)
more
sgpaosacsmC1.a1:
  • Mass concentration of total organics, ambient aerosol in air(total_organics)
  • flow rate measured by the mass flow controller(flow_rate)
  • Inlet closed(inlet_closed)
  • Mass concentration of chloride, ambient aerosol in air.(chloride)
  • Inlet open(inlet_open)
  • Emission current(emission_current)
  • Mass concentration of sulfate, ambient aerosol in air.(sulfate)
  • Mass concentration of ammonium, ambient aerosol in air(ammonium)
  • Vaporizer(vaporizer)
  • Mass concentration of nitrate, ambient aerosol in air(nitrate)
more
sgpaosacsmC1.00:
  • (Raw data stream - documentation not supported)
more
sgpaosacsmC1.b1:
  • Organic mass spectral matrix(org_mx)
  • Mass concentration of ammonium, ambient aerosol in air(ammonium)
  • Mass concentration of sulfate, ambient aerosol in air.(sulfate)
  • Mass concentration of nitrate, ambient aerosol in air(nitrate)
  • Mass concentration of total organics, ambient aerosol in air(total_organics)
  • Mass to charge ratios of ion fragments(amus)
  • Mass concentration of chloride, ambient aerosol in air.(chloride)
  • Organic mass spectral error matrix(org_mx_err)
more

Close this window