DQRID : D121003.2
Start DateStart TimeEnd DateEnd Time
08/30/2012205809/20/20121852
more
Subject:
SGP/SIRS/E13 - Reprocessed: Incorrect UIR calibration
DataStreams:sgpsirsE13.b1, sgpsirs20sE13.a0
Description:
For the time range specified the UIR had incorrect calibration coefficients applied.  The 
k1 coefficient applied was 0.25042 when it should have been 0.2558.  This resulted in the 
instrument under reporting up_long_hemsip by approximately 2.1%. 

Reprocessing completed 20121119
Suggestions: 
Use reprocessed data
Measurements:sgpsirsE13.b1:
  • 10 meter Upwelling Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)
  • Upwelling Longwave Hemispheric Irradiance, Pyrgeometer, Minima(up_long_hemisp_min)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
    Standard Deviation(up_long_hemisp_std)
  • Upwelling longwave hemispheric irradiance, pyrgeometer, maxima(up_long_hemisp_max)
  • Instantaneous Upwelling Longwave Hemispheric Irradiance, Pyrgeometer(up_long_netir)
  • Instantaneous upwelling pyrgeometer case thermistor temperature, pyrgeometer(inst_up_long_case_temp)
  • Instantaneous Dome Temperature(inst_up_long_dome_temp)
more
sgpsirs20sE13.a0:
  • Instantaneous Upwelling Pyrgeometer Dome Thermistor Resistance Pyrgeometer(inst_up_long_dome_resist)
  • Instantaneous Upwelling Shortwave Hemispheric Irradiance, Pyranometer(inst_up_short_hemisp)
  • Instantaneous Upwelling Pyrgeometer Thermopile Pyrgeometer(inst_up_long_hemisp_tp)
  • Instantaneous Upwelling Pyrgeometer Case Thermistor Resistance, Pyrgeometer(inst_up_long_case_resist)
more