DQRID : D121025.2
Start DateStart TimeEnd DateEnd Time
08/14/2012160009/26/20121839
more
Subject:
SGP/SIRS/E33 - Reprocessed: Incorrect UIR calibration
DataStreams:sgpsirsE33.b1, sgpsirs20sE33.a0
Description:
During troubleshooting of the UIR (up_long_hemisp) the UIR and DIR (down_long_hemsip) 
pyrgeometers were swapped to help isolate an instrument noise issue.  

The replacement UIR instrument installed on 8/14 had incorrect calibration coefficients 
applied.  The serial number was correct.

wrong UIR k1 cal = 0.2376
correct UIR k1 cal = 0.2451

Reprocessing completed 20121217.
Suggestions: 
Use reprocessed data.
Measurements:sgpsirsE33.b1:
  • Upwelling Longwave Hemispheric Irradiance, Pyrgeometer, Minima(up_long_hemisp_min)
  • Instantaneous Upwelling Longwave Hemispheric Irradiance, Pyrgeometer(up_long_netir)
  • Upwelling longwave hemispheric irradiance, pyrgeometer, maxima(up_long_hemisp_max)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
    Standard Deviation(up_long_hemisp_std)
  • 10 meter Upwelling Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)
more
sgpsirs20sE33.a0:
  • Instantaneous Upwelling Pyrgeometer Thermopile Pyrgeometer(inst_up_long_hemisp_tp)
  • Instantaneous Upwelling Shortwave Hemispheric Irradiance, Pyranometer(inst_up_short_hemisp)
  • Instantaneous Upwelling Pyrgeometer Dome Thermistor Resistance Pyrgeometer(inst_up_long_dome_resist)
  • Instantaneous Upwelling Pyrgeometer Case Thermistor Resistance, Pyrgeometer(inst_up_long_case_resist)
more