Subject:
PGH/GNDRAD/M1 - Reprocessed: Incorrect calibrations
DataStreams:pghgndrad20sM1.a0, pghgndrad60sM1.b1
Description:
During the entire India deployment the Gndrad system was configured with incorrect 
calibration coefficients for the shortwave instruments. The program hadn't been updated since 
the previous deployment in the Azores.

The instrument mentor, site personal and OSS support have attempted to best determine the 
serial numbers of instruments installed at PGH during the deployment period.  The 
following instrument data will need to be reprocessed with the following correction multipliers:
PIR-UIR(up_long_hemisp): OK, no correction required
PSP-US (up_short_hemsip): 117.97(correct)/129.42(incorrect) = 0.91153
IRT: calibration should be validated by instrument mentor (Vic Morris): mult=25, 
offset=223.2

Correct serial numbers:
PSP-US (up_short_hemsip): 33941F3

Reprocessing completed 20130115
Suggestions: 
Use reprocessed data.
Measurements:pghgndrad60sM1.b1:
  • 25 meter Upwelling Shortwave Hemispheric Irradiance, Pyranometer(up_short_hemisp)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Maxima(up_short_hemisp_max)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer,
    Standard Deviation(up_short_hemisp_std)
  • Upwelling shortwave hemispheric irradiance, pyranometer, minima(up_short_hemisp_min)
  • Surface infra-red temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature, Maxima(sfc_ir_temp_max)
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)
  • Surface Infra-Red Temperature, Minima(sfc_ir_temp_min)
more
pghgndrad20sM1.a0:
  • Instantaneous Upwelling Shortwave Hemispheric Irradiance, Pyranometer(inst_up_short_hemisp)
  • Instantaneous Surface Infra-Red Signal(inst_sfc_ir_signal)
more

Close this window