DQRID : D130108.1
Start DateStart TimeEnd DateEnd Time
12/18/2012150012/21/20122130
more
Subject:
SGP/ACSM/C1 - Pump failure
DataStreams:sgpaosacsmC1.00, sgpaosacsmC1.b1, sgpaosacsmC1.01, sgpaosacsmC1.a1
Description:
The carbon vane vacuum pump that pulls air through the instrument failed again and a 
section of tubing between the ACSM and pump was partially collapsed.  The pump and tubing were 
replaced and returned to service at 2223 GMT on 12/18/2012. The ACSM was then tuned 
before resuming data collection at 2130 GMT on 12/21/2012.
Suggestions: 
Do not use data
Measurements:sgpaosacsmC1.01:
  • (Raw data stream - documentation not supported)
more
sgpaosacsmC1.a1:
  • Mass concentration of total organics, ambient aerosol in air(total_organics)
  • flow rate measured by the mass flow controller(flow_rate)
  • Inlet closed(inlet_closed)
  • Mass concentration of chloride, ambient aerosol in air.(chloride)
  • Inlet open(inlet_open)
  • Emission current(emission_current)
  • Mass concentration of sulfate, ambient aerosol in air.(sulfate)
  • Mass concentration of ammonium, ambient aerosol in air(ammonium)
  • Vaporizer(vaporizer)
  • Mass concentration of nitrate, ambient aerosol in air(nitrate)
more
sgpaosacsmC1.00:
  • (Raw data stream - documentation not supported)
more
sgpaosacsmC1.b1:
  • Organic mass spectral matrix(org_mx)
  • Mass concentration of ammonium, ambient aerosol in air(ammonium)
  • Mass concentration of sulfate, ambient aerosol in air.(sulfate)
  • Mass concentration of nitrate, ambient aerosol in air(nitrate)
  • Mass concentration of total organics, ambient aerosol in air(total_organics)
  • Mass to charge ratios of ion fragments(amus)
  • Mass concentration of chloride, ambient aerosol in air.(chloride)
  • Organic mass spectral error matrix(org_mx_err)
more

Close this window