Subject: | MAG/MWR/M1 -Incorrect retrievals |
DataStreams: | magmwrlosM1.b1, magmwrtipM1.a1
|
Description: | The PWV and LWP retrievals have higher uncertainty because of higher uncertainty in the
calibration due to the ship movement and due to higher uncertainty in the retrieval
coefficients. |
Suggestions: | |
Measurements: | magmwrlosM1.b1: - Averaged total liquid water along LOS path(liq)
- Total water vapor along LOS path(vap)
more
magmwrtipM1.a1: - Total water vapor along zenith path using tip-derived brightness temperatures(vaptip)
- Total liquid water along zenith path using tip-derived brightness temperatures(liqtip)