DQRID : D130729.1
Start DateStart TimeEnd DateEnd Time
05/24/2012000004/24/20130000
more
Subject:
SGP/AOS/C1 - Reprocessed: Bad PSAP flow cal
DataStreams:sgpnoaaaosC1.b1
Description:
There was a leak in the flow calibration device that resulted in a lower than expected 
flow calibration. The data were reprocessed with a flow calibration that is the average of 
the flow cals taken before and after this time period.
Suggestions: 
Reprocessed November 2013
Measurements:sgpnoaaaosC1.b1:
  • Absorption coefficient, blue wavelength, 3 wavelength PSAP, low RH, 1 um size
    cut(Ba_B_Dry_1um_PSAP3W_1)
  • Absorption coefficient, green wavelength, 3 wavelength PSAP, low RH, 10 um size
    cut(Ba_G_Dry_10um_PSAP3W_1)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 10 um size cut(Ba_R_Dry_10um_PSAP3W_1)
  • Absorption Coefficient Red, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_R_Dry_1um_PSAP3W_1)
  • Absorption coefficient, blue wavelength, 3 wavelength PSAP, low RH, 10 um size
    cut(Ba_B_Dry_10um_PSAP3W_1)
  • Absorption coefficient green, corrected for flow and sample area, 3 wavelength
    PSAP, 1 um size cut(Ba_G_Dry_1um_PSAP3W_1)
more