ENA/SKYRAD/C1 - Reprocessed: IRT configuration
DataStreams:enaskyrad60sC1.b1
Description:
SKYRAD IRT sky temperature measurements were ~45K too low. The analog output configuration 
of the IRT was being automatically set to incorrect values by the serial collection 
script due to an error in the name of the startup file.

The data were reprocessed in February 2014 to apply: 
   T_new = (25/26)(T_old-173.2K)+213.2K
Suggestions: 
Retrieve the reprocessed data.
Measurements:enaskyrad60sC1.b1:
  • Sky Infra-Red Temperature Minima(sky_ir_temp_min)
  • Sky infrared temperature, standard deviation(sky_ir_temp_std)
  • Sky Infra Red Temperature Maxima(sky_ir_temp_max)