DQRID : D140318.1
Start DateStart TimeEnd DateEnd Time
02/10/2014000002/13/20141400
more
Subject:
TMP/MWR3C/M1 - Calibration not updated
DataStreams:tmpmwr3cM1.b1
Description:
The radiometer didn't update the calibration because the relative humidity was too high. 
The relative humidity threshold was increased and the radiometer updated its calibration.
Suggestions: 
Measurements:tmpmwr3cM1.b1:
  • Liquid water path(lwp)
  • Mean 23.8 GHz sky brightness temperature(tbsky23)
  • Mean 31.4 GHz sky brightness temperature(tbsky31)
  • Precipitable water vapor(pwv)
more