Subject: | SGP/RL/C1 - Reduced SNR due to partial laser blockage |
DataStreams: | sgprlC1.a0
|
Description: | The laser output was partial blocked. This resulted in reduced SNR in all detection
channels. |
Suggestions: | The data are useable but one may notice a slight increase in the noise. |
Measurements: | sgprlC1.a0: - Summed analog signal in the temperature 2 channel(t2_analog_high)
- Number of photons counted in the low water vapor channel(water_counts_low)
- Summed analog signal in the low water vapor channel(water_analog_low)
- Summed analog signal in the high depolarization channel(depolarization_analog_high)
- Number of photons counted in the high nitrogen channel(nitrogen_counts_high)
- Summed analog signal in the liquid water channel(liquid_analog_high)
- Number of photons counted in the temperature 2 channel(t2_counts_high)
- Count rate in the high depolarization channel(depolarization_counts_high)
- Number of photons counted in the high elastic (aerosol) channel(elastic_counts_high)
- Count rate in the high water vapor channel(water_counts_high)
- Number of photons counted in the temperature 1 channel(t1_counts_high)
- Summed analog signal in the low nitrogen channel(nitrogen_analog_low)
- Summed analog signal in the high water vapor channel(water_analog_high)
- Summed analog signal in the high elastic channel(elastic_analog_high)
- Summed analog signal in the high nitrogen channel(nitrogen_analog_high)
- Number of photons counted in the liquid water channel(liquid_counts_high)
- Number of photons counted in the low elastic (aerosol) channel(elastic_counts_low)
- Summed analog signal in the low elastic channel(elastic_analog_low)
- Summed analog signal in the temperature 1 channel(t1_analog_high)
- Number of photons counted in the low nitrogen channel(nitrogen_counts_low)
more
|