Subject:
TWP/ECOR/E31 - Reprocessing Notice (RID 1121)
DataStreams:twp30ecorE31.b1
Description:
See below for details of reprocessing that was completed on 20141028.  Original DQR 
category: 4 (Red, Incorrect).

-----------------------------------------------------------

The Mean Water Vapor Density and ground analog signal leads were switched.

This resulted in:
1) The sign of Latent Heat Flux and Mean Water Vapor Density being reversed
2) Mean CO2 density concentration being referenced to the Mean Water Vapor Density signal 
instead of to ground.