Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/29/2014 | 0234 | 05/02/2014 | 2127 |
Subject: | MAO/AOS/S1 - Impactor state incorrect |
DataStreams: | maoaosimpactorS1.a1 |
Description: | Impactor state was flatline at 0. During this period, the 'Master' computer went into a state with constant reboots. The Master controls and records the impactor state. From the flow record of the Nephelometer and the PSAP it is apparent the impactor did not cycle during this period. |
Suggestions: | null |
Measurements: | maoaosimpactorS1.a1: more
|